TOC-4200 - Specs
On-line TOC Analyzer
TOC-4200
| Measurement items | 
                                   TC, NPOC  | 
                              
|---|---|
| Measurement principle | 
                                   680℃ combustion catalytic oxidation - NDIR detection method  | 
                              
| Measurement range | 
                                   From 0 - 5 to 0 - 1,000 mgC/L f.s. (0 to 20,000 mgC/L f.s. with dilution function)  | 
                              
| Repeatability | 
                                   Within ±2 % f.s.*1  | 
                              
| Span stability | 
                                   Within ±2 % f.s. /day (temperature variation within 5℃)*1  | 
                              
| Zero stability | 
                                   Within ±2 % f.s. /day (temperature variation within 5℃)*1  | 
                              
| Measurement cycle | 
                                   4 minutes min. (NPOC, residual IC 2 % max.)*2  | 
                              
| Sample injection method | 
                                   Syringe pump/sliding injection port  | 
                              
| IC removal method | 
                                   Acidify/sparge within syringe  | 
                              
| Sample dilution function | 
                                   Dilution within syringe; dilution factor 2 to 50  | 
                              
| Automatic calibration function | 
                                   Automatic calibration using standard solution (Dilution water is used as the zero standard solution for zero-calibration.)   | 
                              
| Stream switching function | 
                                   Streams 1 to 6 can be switched (with the relevant option)  | 
                              
| Display, operation | 
                                   Color LCD with touch screen  | 
                              
| Conversion function | 
                                   Conversion with a linear expression using required indices (Conversion expression must be determined in advance.)  | 
                              
| Load calculation function | 
                                   Using the analog input board (option) permits load calculations. (Non-isolated analog inputs; load resistance: 100 ohms)  | 
                              
| EPA TOC removal rate calculation function | 
                                   TOC removal rate calculation according to the United States EPA regulations (Part IV 40 CFR Part 9, 141 and 142, 1998) (when a 2- or more-stream switching option is provided)  | 
                              
| Data storage function | 
                                   Store 20,000 past on-line measured values (equivalent to one year's data at 30-minute measuring cycle).  | 
                              
| Data storage device | 
                                   Store measured values or measurement conditions to USB memory*3.  | 
                              
| Recorder | 
                                   Recording width 100 mm, 6-dot recorder (optional)  | 
                              
| Printer | 
                                   Thermal printer, 40 characters per line; 110 mm chart width (optional)  | 
                              
| Analog output | 
                                   Selectable between 4 – 20 mA DC or 0 – 16 mA DC (isolated outputs)  | 
                              
| Contact output signal | 
                                   Alarm signals  | 
                              
| Contact input | 
                                   Start calibration, stop measurement, start on-line measurement, reset alarm, measurement start flow line 1 to 6, halt measurement (output contact capacity: 1.4 V DC min., 10 mA min.)  | 
                              
| Communication function | 
                                   Compatible with RS-485, RS-232C (option), and Modbus  | 
                              
| Carrier gas | 
                                   Pressurized air, high-purity air, oxygen*4; supply pressure: 250 to 300 kPa  | 
                              
| Power supply | 
                                   AC 100 to 240 V ±10%, 10 A, 50/60 Hz  | 
                              
| Ambient temperature | 
                                   Within 1 to 40℃  | 
                              
| Construction | 
                                   Indoor wall mounting (optional stand available)  | 
                              
| Dimensions | 
                                   W550 x D383 x H1240 mm (excluding protrusions)  | 
                              
| Weight | 
                                   Approx. 70 kg (154 lbs.)  | 
                              
*1) With automatic settings
*2) When measured sample is potassium hydrogen phthalate solution and operations to prepare for the next measurement are performed.
*3) If TN option is installed, oxygen and high-purity nitorogen cannot be used.
TOC-4200
| Measurement principle | 
                                   680°C combustion catalytic oxidation  | 
                              
|---|---|
| Measurement Items (*Optional) | 
                                   NPOC, TC, TOC (TC-IC)*  | 
                              
| Measurement range | 
                                   0-5 to 0-1,000 mgC/Lf.s.  | 
                              
| Measurement cycle | 
                                   4 minuts min. (Using NPOC)  | 
                              
| Repeatability | 
                                   Within ±2% f.s.*  |